NMR Data Acquisition
Collect Digital Data
    ADC – analog to digital converter
SR = 1 / (2 * SW)
The Nyquist Theorem says that we have to sample at least twice as fast as the fastest (higher frequency) signal.
Sample Rate
- Correct rate, correct frequency
-½ correct rate, ½ correct frequency Folded peaks!
Wrong phase!
SR – sampling rate